03 / Proper protocol
A reliable review follows the entire test—not one printout.
Florida's operational forms are model-specific. The outline below captures the core current safeguards, but the actual approved form and rule in effect on the test date control. Florida applies a substantial-compliance standard, so a deviation does not automatically equal exclusion.
Qualified operator and approved instrument
Confirm the operator's permit, instrument make and model, serial number, registration, approved software, and whether the instrument was in evidentiary service.
Observation before sampling
Florida's current operational procedures use a minimum 20-minute observation period. The evidence should show when it began and what occurred during it.
Blanks, diagnostics, and controls
The sequence checks for a clean air path, successful instrument diagnostics, and an acceptable result from a known alcohol standard before relying on a subject result.
Two valid subject samples
The instrument seeks separate samples. Under the current 9000 procedure, a third is requested automatically when the first two do not agree within 0.020 g/210L.
Post-test safeguards
Model-specific procedures run additional control, blank, or diagnostic checks after the subject samples. Review their values and any status messages.
Preserved result and affidavit
Compare the printout or affidavit with electronic data, timestamps, video, operator testimony, inspection files, repair entries, and nearby tests on the same instrument.